• CT值均勻性評價方法研究
    中國測試楊克檉
    摘  要:針對CT值均勻性傳統定義只考慮圖像的中心和外圍部分而不能全面反映整個視野(FOV)內CT值變化的缺點,提出一種改進的評價方法,以適應CT新技術質量保證檢測的需要。采用直徑為20 cm的圓柱形水模,在GE Discovery CT750 HD掃描儀上選擇常用頭部協議進行軸向掃描。在水模圖像的3,6,9,12點鐘方向上距離中心0,4.0,8.0 cm處取面積為1 cm2的圓形感興趣區,分別計算CT值均值和標準差。定義其中最大、最小均值之差為CT值均勻性指數,最大標準差為噪聲指數;單能圖像序列中CT值均勻性指數最大值定義為該序列的CT值均勻性指數,噪聲指數最大值為該序列噪聲指數。結果表明:從FOV中心向外CT值并非單調變化,CT值均勻性評價新方法更加全面、客觀地評價整個FOV的圖像質量,并能反映雙能CT的不同探測器陣列對應圖像之間以及不同能級圖像序列之間的質量差異。
    關鍵詞:計算機體層攝影;雙能;CT值均勻性;質量保證
    文獻標志碼:A       文章編號:1674-5124(2016)01-0007-05
    A modified evaluation method for CT value uniformity
    YANG Kecheng
    (School of Biomedical Engineering,Third Military Medical University,Chongqing 400038,China)
    Abstract: In current quality assurance (QA) testing guidelines and standards, CT value uniformity (CTVU) is defined only based on image centers and peripheries; instead, the variation in CT value can not be fully reflected within the whole field of view. To solve this problem, a modified method was proposed to meet the QA testing needs of new CT scanners. A 20 cm-diameter water equivalent plastic cylindrical phantom filled with pure water was scanned in a step-and-shot way with a routine dual-energy head protocol (GSI-26) on a GE Discovery CT750 HD scanner. Images were analyzed with a customer designed MATLAB program. Nine circular ROIs (1 cm2) were selected at the spots that were 0 cm, 4.0 cm and 8.0 cm away from the centroid at 3, 6, 9, 12 o’clock positions to calculate the CT mean value and standard deviations. The differences between the maximum and minimum mean value were defined as CTVU indexes; the maximal SD was defined as noise figure; the maximal CTVU index and noise index in virtual monochromatic image sequence were defined as CTVU index and noise index respectively. Experimental results show that the modified evaluation method reflects not only the CT value variation within the whole field of view in a comprehensive and objective manner, but also the quality differences between different images scanned by different detector slices and virtual monochromatic image sequences of dual-energy CT.
    Keywords: computed tomography(CT); dual energy; CT value uniformity; quality assurance
     
     
    網站首頁  |  關于我們  |  聯系我們  |  廣告服務  |  版權隱私  |  友情鏈接  |  站點導航
     
    九九久久精品国产| 国内精品视频在线观看| 亚洲youwu永久无码精品| 亚洲国产精品久久网午夜| 久久精品国产亚洲av麻豆| 国产揄拍国产精品| 99re国产精品视频首页| 精品国产高清在线拍| 国产乱子伦精品无码专区| 亚洲精品视频在线看| 日韩美女18网站久久精品| 国产在线精品二区韩国演艺界| 日韩国产精品99久久久久久| 国产精品久久久久久精品三级| 麻豆精品人妻一区二区三区蜜桃| 亚洲精品无码不卡在线播放| 亚洲精品美女久久久久久久| 久久久国产精品va麻豆 | 538prom精品视频线放| 三上悠亚日韩精品一区在线| 精品午夜福利1000在线观看| 9久热这里只有精品免费| 91久久婷婷国产综合精品青草| 国产亚洲精品看片在线观看| 中国精品18videosex性中国| 精品综合久久久久久97超人| 久久精品视频观看| 999久久久免费精品播放| 精品国产一区二区三区久久久狼 | 国产成人综合精品一区| 精品91自产拍在线观看二区| 国产69精品久久久久久人妻精品| 九九热在线精品视频| 99视频精品全部免费观看| 中文字幕动漫精品专区| 国产乱码精品一区二区三区麻豆 | 无码人妻精品一区二区蜜桃网站 | 无码精品A∨在线观看| 亚洲国产精品专区| 成人啪精品视频免费网站| 综合91在线精品|